Three FDA-cleared a2z Radiology AI devices now cover 26 abdominopelvic CT findings. No published sensitivity or specificity exists for any of them. The RSNA data hospitals will see this week comes from a different set of products entirely.

What 26 Findings Across Three Devices Actually Means
On November 19, 2026, a2z Radiology AI published a press release announcing that the FDA had cleared its third device for abdominopelvic CT triage. The new product, a2z-Omni-Triage, can flag nine additional findings: moderate or large layering pleural effusion, pneumothorax, complex pleural collection, bowel perforation, moderate or severe hydronephrosis, large aortic aneurysm (5 cm or larger), non-sacral pelvic fracture, sacral or coccygeal fracture, and soft tissue abscess.
That brings a2z's total to 26 FDA-cleared triage findings across three devices.
The first device, a2z-Unified-Triage, cleared in November 2025, flags seven conditions: small bowel obstruction, acute cholecystitis, acute pancreatitis, acute diverticulitis, hydronephrosis, free air, and unruptured abdominal aortic aneurysm. The second, a2z-Abdo-Triage, cleared in September 2026, adds 10 more: obstructing ureteral stone, hydroureter, retroperitoneal hematoma, large volume ascites, acute appendicitis, intraabdominopelvic abscess, hemoperitoneum or complex free fluid, splenic traumatic injury, hepatic traumatic injury, and renal traumatic injury. That second device also carries FDA Breakthrough Device designation, which is a separate designation from clearance, granted to expedite development for serious conditions.
The third device, Omni-Triage, extends coverage to chest-adjacent findings when that anatomy appears in the abdominopelvic scan.
Each device is cleared as an adjunctive tool. The indication is consistent across all three: analyze abdominopelvic CT scans acquired with or without IV contrast in adults aged 22 and older, flag suspected positive cases, and communicate results to help radiology teams prioritize studies. Radiologists remain responsible for reviewing the full scan and making clinical judgments.
The company's co-founders are Pranav Rajpurkar, PhD, an associate professor of biomedical informatics at Harvard Medical School, and Samir Rajpurkar, CEO. The firm is backed by Khosla Ventures and SeaX Ventures.
Abdomen-pelvis CT is, by the company's own account, the highest-volume CT category in the U.S., with more than 20 million exams annually. The clinical argument for triage AI in that space is real: a radiologist working a long overnight shift faces a worklist in which a scan ordered as routine can carry an urgent finding that gets reviewed last. The question is not whether the problem is worth solving. It is whether anyone has published the evidence that these specific products solve it.

The Performance Numbers That Were Never Published
Here is the central fact about all three a2z-cleared devices: no published performance data exists for any of them.
Not "limited data." Not "preliminary data." No sensitivity, specificity, or area under the receiver operating characteristic curve is publicly available for a2z-Unified-Triage, a2z-Abdo-Triage, or a2z-Omni-Triage.
The November 2026 press release announcing the third clearance is consistent with the two before it. All three describe the clinical use case, list the cleared findings, name the hospital IT workflow that the product fits into, and provide executive quotes. None publishes the performance testing that the FDA evaluated in clearing the device.
This is worth clarifying, because the company's press language implies rigor without disclosing it. The November 2026 release says the nine new findings "were cleared with their own definitions and performance testing." That sentence confirms that testing existed. It does not say what the results were, or what populations the tests used, or what the prevalence of each finding was in the testing dataset, or whether the test population resembles the range of patients hospitals actually see.
The FDA's 510(k) database does contain performance summaries. For a2z-Omni-Triage, a 510(k) number appears in regulatory coverage. But the full performance summary, with the actual numbers, is not in the company's press materials and was not in any published peer-reviewed article as of this writing.
That gap matters for a specific reason. Sensitivity on a testing dataset assembled by the manufacturer, under controlled conditions, often differs from sensitivity on the worklist your hospital actually runs. That difference is not a flaw unique to a2z. It is structural. It is how FDA clearance works for most AI medical devices.

RSNA 2026 Research Data: Impressive, With a Critical Caveat
This week, the Radiological Society of North America meets in Chicago for RSNA 2026. Four studies from Mass General Brigham researchers are being presented there, all involving a2z models. The numbers are worth reading carefully, because they are real, they are from a credible institution, and they describe a product that is not what hospitals can actually buy.
The four studies use a2z "research models," which are explicitly research use only and not cleared for clinical or commercial use. This distinction matters because the same company name appears in both the product marketing and the research presentations, but the products are different.
What the research models showed at MGB:
For 962 consecutive inpatient abdominopelvic CT exams at two academic hospitals, the research model identified 88.9% of high-priority findings, with the two hospitals within four percentage points of each other.
On paired low-dose and standard-dose abdominal scans, model outputs agreed 95.7% of the time, suggesting the research model is reasonably stable across dose variation.
In a draft-report study being presented as an oral on December 2, 96.4% of AI-drafted reports for seven target conditions were correct. 2.7% of drafted findings needed editing. The model included 70.8% of expected positive findings. On a 300-study radiologist re-read, the model surfaced findings absent from the original reports that were confirmed on images, including a pulmonary embolism.
For a pediatric setting, an adult-trained research model reached an average AUC of 0.964 on five pediatric emergencies (small bowel obstruction, appendicitis, hydronephrosis, liver trauma, spleen trauma) without retraining or threshold adjustments.
Those are strong numbers for a research system. The 88.9% inpatient detection rate, the AUC of 0.964 in pediatric cases the model was not trained on, the pulmonary embolism catch in a re-read study: all of these are findings a hospital administrator would want to put in a procurement document.
The problem is they cannot, accurately. These are not the cleared products. They are research models. A hospital buying a2z-Omni-Triage is not buying the model that showed AUC 0.964 on pediatric scans. It is buying a product whose performance characteristics are, at the moment of this writing, not publicly stated.
Mannudeep K. Kalra, MD, director of the Webster Center for Quality and Safety at MGB and a professor of radiology at Harvard Medical School, is quoted in the company's RSNA announcement: "The important question was not whether an AI model could find one condition." That framing is correct. The more urgent question is whether the cleared product performed like the research model, and in what populations.

What the Broader CT Triage AI Evidence Record Shows
a2z is not an outlier. The pattern of FDA-cleared AI devices with thin public evidence is a documented feature of the current market, not an exception.
A PLOS Digital Health study published August 19, 2026, by researchers at MIT, examined 1,357 FDA AI- and machine-learning-enabled devices cleared or approved through December 5, 2025. Of those 1,357 devices: 34, or 2.5%, were linked to a registered prospective trial. Only 3, or 0.2%, were evaluated on patient-centered outcomes such as mortality, morbidity, or readmission. Twelve devices, or 0.9%, had peer-reviewed publications.
Translated: 97.5% of AI devices entered clinical use with no registered prospective trial behind them.
For radiology specifically, the numbers are steeper. More than three-quarters of FDA AI devices are used in radiology. Of 1,059 radiology-specific AI devices in the dataset, only 3 had a registered trial. That is 0.3%.
A co-author of the study, Ordonez, was quoted in AuntMinnie: "Clearance is the beginning of due diligence, not the end."
The 34 trials that did exist had their own limitations. 94% were run by the device manufacturer. Nearly three-quarters enrolled fewer than 500 patients. 68% ran only in the U.S. Only 27% reported any subgroup analysis. Race or ethnicity was reported in only 3 of 34 trials. Pediatric patients were excluded from almost all of them.
Aidoc, the other major CT triage AI company, also expanded its FDA clearances in 2026. In January 2026, Aidoc received clearance for a CT-based platform with 14 cleared indications. The company reported 97% mean sensitivity and 98% mean specificity across the 11 new indications. Those are vendor-reported figures. No sample size, no study design, and no clearance number appeared in coverage of the announcement.
The University of Washington presented real-world deployment data at RSNA 2026 that illustrates the range of outcomes. Researchers at a Level 1 trauma center deployed 11 triage AI algorithms across three hospitals from September 2023 through August 2025. The systems processed more than 742,000 imaging exams. The results varied substantially.
For endotracheal tube malposition, median report turnaround time dropped about 76%. For rib fracture, it dropped about 53%. For pulmonary embolism and intracranial hemorrhage, there was no improvement. The researchers attributed the flat results for PE and ICH to higher proportions of follow-up or inpatient studies with known findings, in which triage prioritization adds no value because the finding is already expected.
The takeaway from the UW data is not that CT triage AI fails. It is that outcomes vary by application, and that realizing benefit requires knowing which applications are likely to perform before deployment. That requires data the manufacturers are not currently publishing.

How the FDA's Clearance Standard Works Here
The FDA's 510(k) pathway, which governs most AI medical devices, requires that a manufacturer demonstrate substantial equivalence to a device already on the market. It does not require proof of patient benefit. It does not require a randomized controlled trial. It does not require the manufacturer to publish the performance testing that was submitted.
A citizen petition filed in 2025, FDA-2025-P-5560, proposed exempting radiology CADe/CADx and computer-aided triage devices from premarket notification requirements entirely. The FDA denied that petition on April 1, 2026. That denial preserves the 510(k) requirement. It does not change what the 510(k) requires.
The MIT team's recommended fix is straightforward: mandatory pre-registration of prospective trials for Class II and III AI devices before clearance, followed by multicenter outcome trials of at least 2,000 patients with pre-specified subgroup analyses after clearance. Journals could enforce SPIRIT-AI and CONSORT-AI reporting standards. Payers could tie reimbursement to demonstrated benefit.
None of those requirements currently apply to a2z's cleared products.
This is not an argument that the products are unsafe. The FDA's performance testing does something important: it establishes that the device doesn't flag findings at a rate that would be harmful. What it does not establish, and what the current evidence record does not establish, is the magnitude of clinical benefit in the populations a hospital actually serves.
The question a radiologist at a medium-volume community hospital should be asking is not whether the system cleared FDA. It is: in what population did it perform, at what prevalence, and does that match my worklist? The answer to that question requires public data that does not currently exist.

What Hospitals Should Ask Before Signing a Contract
The case for CT triage AI is legitimate. Emergency radiology worklists are long. High-acuity findings get missed. Any tool that reliably flags a bowel perforation or a large aortic aneurysm on a scan that would otherwise wait until morning has real value.
The commercial pitch for 26 cleared findings has a parallel logic: more coverage is better, and each additional finding is another way the system earns its place in the workflow. That logic holds if the performance is consistent across findings. The problem is that the performance for each of the 26 findings was tested separately, under separate definitions, in separate datasets. "26 cleared findings" does not mean 26 findings all performing at the same sensitivity threshold. It means 26 separate regulatory submissions, each evaluated independently.
Before a hospital procurement team signs a contract for any multi-finding CT triage AI, including a2z's products, there are specific questions worth asking:
What is the published sensitivity and specificity for each finding, and in what population? Company executives talking about research model data at RSNA is not the same as published clearance-device performance.
What prevalence was used in the performance testing for each finding? A finding that affects 1 in 100 CT scans behaves differently in a triage algorithm than one that affects 1 in 1,000. At low prevalence, even high sensitivity produces many false alarms. The MIT team calculated that at 1% prevalence, a 97% accurate alert is right only 28% of the time.
What adult age floor was tested? All three a2z devices specify adults aged 22 and older. What happens at 18, 19, 20, 21? What happens to findings that present differently in patients who are 70 versus 30?
What is the alert workflow and who gets notified? The system flags and communicates suspected positive cases to help teams prioritize. How that notification reaches the radiologist on call, what happens if it is missed, and who carries liability when an alert fires and is not acted on are questions that go beyond the product and into the hospital's own workflow.
Is the RSNA research data from the cleared product? The answer for a2z is currently: no. The four MGB studies presented at RSNA 2026 this week used research models. Hospitals evaluating a2z products should ask for the performance data specific to a2z-Unified-Triage, a2z-Abdo-Triage, and a2z-Omni-Triage as cleared, not for the research model data.
What postmarket surveillance is the company committed to? Real-world performance can diverge from testing performance as the worklist mix changes. A vendor with no commitment to publishing postmarket data is a vendor with no obligation to tell you when performance degrades.
Is the finding coverage complementary to what you already have? Aidoc, Aidoc's competitors, and a2z now have overlapping cleared indications. Pulmonary embolism, intracranial hemorrhage, pneumothorax, and large vessel findings may already be covered by an existing deployment. Adding a system that re-covers ground you already have is not 26 new findings: it is some subset of 26, and the incremental value should be evaluated on that basis.
Connecting to the Screening Gap
The lung cancer screening literature contains a lesson that applies directly here. Among the highest-acuity findings on a2z's cleared list is unruptured abdominal aortic aneurysm. Among the new Omni-Triage findings is large aortic aneurysm (5 cm or larger). Those are findings that benefit from early detection. The clinical evidence for screening those conditions is established. The clinical evidence for using AI triage to accelerate detection of those conditions, in real-world settings, is not.
Oatmeal Health works at the intersection of prevention, screening, and access. What we see repeatedly in the lung cancer screening data applies to CT triage AI as well: the technology exists, the regulatory pathway has been completed, and the evidence for how well the product actually performs in the populations who need it most is sparse. For lung cancer, the gap is between high-quality national data and the community practice settings where most patients live. For CT triage AI, the gap is between the testing population that earned the clearance and the real-world worklist in a Level 2 community hospital at 2 a.m.
Closing that gap does not require waiting. It requires that vendors publish the performance data they already have, that health systems ask for it before signing contracts, and that RSNA presentations distinguish clearly between what research models showed and what cleared products are built to do.
The 26-finding milestone is real. Whether those findings are detected reliably in your hospital's patient population is a question the press release does not answer.
Jonathan Govette is Co-Founder and CEO of Oatmeal Health, a company focused on improving access to lung cancer screening and early detection for underserved populations. This newsletter covers healthcare technology, payment policy, and the evidence gaps between regulatory clearance and clinical practice.